AITR Series Medical Insulation Transformer

  • Insulation Transformer for Hospital IT Power System

  • Double Insulation Treatment

  • Electrostatic Shielding Layer

  • PT100 Temperature Sensor Embedded in the Winding

  • CE; IEC 61558-1:2009; IEC 61558-2-15:2011; IEC60364-7-710:2016

Features of AITR Series Medical Insulation Transformer

Insulation Transformer for Hospital IT Power System

● Inrush Current(IE): <12Ln

● Short-circuits Voltage: <3% Un

● Noise Level: <40dB (A)

● Efficiency: >96%

Insulation Transformer for Hospital IT Power System

Multiple Insulation Treatment

● Leakage Current:  <200μA

● Insulation Class: H


Multiple Insulation Treatment

PT100 Temperature Sensor Embedded in the Winding

● Thermistor: PT100

● Temperature Measuring Range: -50℃~+200℃

PT100 Temperature Sensor Embedded in the Winding


Specifications of AITR Series Medical Insulation Transformer

Rated Capacity
3.15kVA/5kVA/6.3kVA/8kVA/10kVA
Frequency50/60Hz
Rated input voltage230V
The output voltage230V/115V
Inrush current<12In
Leakage Current<200uA
No-load output voltage<235V
No-load output current<3% In
Short-circuit voltage<3% Un
Efficiency>96%
Maximum ambient temperature<40℃
No-load temperature rise<33℃
Full load temperature rise<76℃
Withstand voltage4200V/Min
Insulation ClassH
Noise Level<40dB (A)