E-mail Us
operating room electrical panels
operating room electrical panels
isolation panel for operating rooms
hospital isolation panel wiring diagram

AIM-M300 Medical Intelligent Insulation Monitor

  • Insulation monitoring for medical IT systems

  • Load and temperature monitoring of isolation transformer

  • Device disconnection monitoring

  • LED indicators: On, Insulation, Overload, Overtemp

  • Alarm relay output

  • Test button

  • Insulation fault location

  • Din 35mm

  • RS485 (MODBUS-RTU) + CAN

Specifications of AIM-M300 Medical Intelligent Insulation Monitor

Auxiliary PowerVoltageDC 18…36V
Power consumption≤6W
Insulation MonitoringResistance15…999kΩ
measuring
range
Response value50…999kΩ
Relative±10%,±10K
uncertainty
Response time≤3s
Permissible≤5uF
system leakage
capacitance Ce
Measuring≤12V
voltage Um

Measuring≤5uA
current Im
Impedance Zi≥200kΩ
Internal DC≥240kΩ
resistance Ri

Permissible≤DC280V
extraneous DC
voltage Ufg
Temperature MonitoringThermal2 Pt100
resistor
Measuring-50…+200℃
range
Alarm Value Range0…+200℃
Current MonitoringAlarm Value5…60A
Measuring
accuracy
±5%


Drawings of AIM-M300 Medical Intelligent Insulation Monitor

Thank you for your attention to Acrel
Let's know your need of smart energy meter.