Let's know your need of smart energy meter.
Acrel will get in touch with you ASAP.
Insulation monitoring for medical IT systems
Load and temperature monitoring of isolation transformer
Device disconnection monitoring
LED indicators: On, Insulation, Overload, Overtemp
Alarm relay output
Test button
Insulation fault location
Din 35mm
RS485 (MODBUS-RTU) + CAN
Auxiliary Power | Voltage | DC 18…36V |
Power consumption | ≤6W | |
Insulation Monitoring | Resistance | 15…999kΩ |
measuring | ||
range | ||
Response value | 50…999kΩ | |
Relative | ±10%,±10K | |
uncertainty | ||
Response time | ≤3s | |
Permissible | ≤5uF | |
system leakage | ||
capacitance Ce | ||
Measuring | ≤12V | |
voltage Um | ||
Measuring | ≤5uA | |
current Im | ||
Impedance Zi | ≥200kΩ | |
Internal DC | ≥240kΩ | |
resistance Ri | ||
Permissible | ≤DC280V | |
extraneous DC | ||
voltage Ufg | ||
Temperature Monitoring | Thermal | 2 Pt100 |
resistor | ||
Measuring | -50…+200℃ | |
range | ||
Alarm Value Range | 0…+200℃ | |
Current Monitoring | Alarm Value | 5…60A |
Measuring accuracy | ±5% | |