AIM-M100 Medical Isolated Meter

  • Insulation monitoring for medical IT systems

  • Load and temperature monitoring of isolation transformer

  • Device disconnection monitoring

  • LED indicators: On, Insulation, Overload, Overtemp

  • Alarm relay output

  • Test button

  • Insulation fault location

  • Din 35mm

  • RS485 (MODBUS-RTU) + CAN

Specifications of AIM-M100 Medical Isolated Meter

AUX PowerVoltageAC220V±10%Temperature monitoringThermal resistor2 Pt100
Power consumption≤5VAMeasuring range-50~+200℃
Insulation MonitoringResistance measuring range10~999kΩAlarm value range0~+200℃
Response value50~999kΩAlarmOutput mode1 Relays
Relative uncertainty±10%outputContact ratingAC 250V/3A

DC 30V/3A
Response time≤3sEnvironmentOperating temperature-10~+55℃
Permissible system leakage capacitance Ce≤5uFTransport temperature-25~+70℃
Measuring voltage Um≤12VStorage temperature-25~+70℃
Measuring current Im≤50uARelative humidity5%-95%, No condensation
Impedance Zi≥200kΩAltitude≤2500m
Internal DC resistance Ri≥240kΩIP degreeIP30
Permissible extraneous DC voltage Ufg≤DC280VRated impulse voltage / pollution degree4KV/Ⅲ
Load Current MonitoringMeasuring Value2.1~50AEMCIEC 61326-2-4
Alarm Value5~50ACommunicationRS485(Modbus-RTU)
Measuring accuracy±5%